600 MHz Cryogenic NMR :A core analytical instrument that leverages cryogenic technology to overcome detection limits and empower cutting-edge research. With cryogenic technology as its key advantage, combined with the 600 MHz high resonance frequency, it effectively suppresses interference from molecular thermal motion, enhances spectral resolution and detection sensitivity, and accurately captures resonance signals of non-zero-spin nuclei. It efficiently analyzes challenging systems such as biomacromolecule conformations, rare nuclides, weak interactions, and material microstructures, and supports in-depth studies of molecular dynamic processes and drug action mechanisms. It is applicable to biomedical science, chemistry, materials science, pharmacy, environmental science, geology, and other fields. Main specifications: Magnetic field strength 14.9 T; room-temperature bore diameter 54 mm; field drift <6 Hz/h; liquid helium hold time ≥365 days; liquid nitrogen hold time ≥16 days; vertical 5-G line <1.4 m, horizontal 5-G line <0.7 m; Z-gradient strength 50 G/cm; time resolution 12.5 ns; frequency resolution <0.005 Hz; phase resolution <0.006°; variable temperature range −150 °C to +200 °C with accuracy <±0.1 °C; detectable nuclei: ¹H, ¹⁹F, and all nuclei with resonance frequencies between ³¹P and ¹⁰⁹Ag.